TY - JOUR
T1 - 3D free-hand ultrasound to register anatomical landmarks at the pelvis and localize the hip joint center in lean and obese individuals
AU - Horsak, Brian
AU - Schwab, Caterine
AU - Durstberger, Sebastian
AU - Thajer, Alexandra
AU - Greber-Platzer, Susanne
AU - Kainz, Hans
AU - Jonkers, Ilse
AU - Kranzl, Andreas
N1 - Publisher Copyright:
© 2021, The Author(s).
PY - 2021/5/20
Y1 - 2021/5/20
N2 - 3D free-hand ultrasound (3DFUS) is becoming increasingly popular to assist clinical gait analysis because it is cost- and time-efficient and does not expose participants to radiation. The aim of this study was to evaluate its reliability in localizing the anterior superior iliac spine (ASIS) at the pelvis and the hip joint centers (HJC). Additionally, we evaluated its accuracy to get a rough estimation of the potential to use of 3DFUS to segment bony surface. This could offer potential to register medical images to motion capture data in future. To evaluate reliability, a test–retest study was conducted in 16 lean and 19 obese individuals. The locations of the ASIS were determined by manual marker placement (MMP), an instrumented pointer technique (IPT), and with 3DFUS. The HJC location was also determined with 3DFUS. To quantify reliability, intraclass correlation coefficients (ICCs), the standard error of measurement (SEm), among other statistical parameters, were calculated for the identified locations between the test and retest. To assess accuracy, the surface of a human plastic pelvic phantom was segmented with 3DFUS in a distilled water bath in 27 trials and compared to a 3D laser scan of the pelvis. Regarding reliability, the MMP, but especially the IPT showed high reliability in lean (SEm: 2–3 mm) and reduced reliability in obese individuals (SEm: 6–15 mm). Compared to MMP and IPT, 3DFUS presented lower reliability in the lean group (SEm: 2–4 mm vs. 2–8 mm, respectively) but slightly better values in the obese group (SEm: 7–11 mm vs. 6–16 mm, respectively). Correlations between test–retest reliability and torso body fat mass (% of body mass) indicated a moderate to strong relationship for MMP and IPT but only a weak correlation for the 3DFUS approach. The water-bath experiments indicated an acceptable level of 3.5 (1.7) mm of accuracy for 3DFUS in segmenting bone surface. Despite some difficulties with single trials, our data give further rise to the idea that 3DFUS could serve as a promising tool in future to inform marker placement and hip joint center location, especially in groups with higher amount of body fat.
AB - 3D free-hand ultrasound (3DFUS) is becoming increasingly popular to assist clinical gait analysis because it is cost- and time-efficient and does not expose participants to radiation. The aim of this study was to evaluate its reliability in localizing the anterior superior iliac spine (ASIS) at the pelvis and the hip joint centers (HJC). Additionally, we evaluated its accuracy to get a rough estimation of the potential to use of 3DFUS to segment bony surface. This could offer potential to register medical images to motion capture data in future. To evaluate reliability, a test–retest study was conducted in 16 lean and 19 obese individuals. The locations of the ASIS were determined by manual marker placement (MMP), an instrumented pointer technique (IPT), and with 3DFUS. The HJC location was also determined with 3DFUS. To quantify reliability, intraclass correlation coefficients (ICCs), the standard error of measurement (SEm), among other statistical parameters, were calculated for the identified locations between the test and retest. To assess accuracy, the surface of a human plastic pelvic phantom was segmented with 3DFUS in a distilled water bath in 27 trials and compared to a 3D laser scan of the pelvis. Regarding reliability, the MMP, but especially the IPT showed high reliability in lean (SEm: 2–3 mm) and reduced reliability in obese individuals (SEm: 6–15 mm). Compared to MMP and IPT, 3DFUS presented lower reliability in the lean group (SEm: 2–4 mm vs. 2–8 mm, respectively) but slightly better values in the obese group (SEm: 7–11 mm vs. 6–16 mm, respectively). Correlations between test–retest reliability and torso body fat mass (% of body mass) indicated a moderate to strong relationship for MMP and IPT but only a weak correlation for the 3DFUS approach. The water-bath experiments indicated an acceptable level of 3.5 (1.7) mm of accuracy for 3DFUS in segmenting bone surface. Despite some difficulties with single trials, our data give further rise to the idea that 3DFUS could serve as a promising tool in future to inform marker placement and hip joint center location, especially in groups with higher amount of body fat.
KW - ACCURACY
KW - BODY-MASS INDEX
KW - CHILDREN
KW - FEMUR
KW - GAIT ANALYSIS
KW - KINEMATICS
KW - MOTION ANALYSIS
KW - RELIABILITY
KW - SOFT-TISSUE
KW - SYSTEM
UR - http://www.scopus.com/inward/record.url?scp=85106274137&partnerID=8YFLogxK
U2 - 10.1038/s41598-021-89763-7
DO - 10.1038/s41598-021-89763-7
M3 - Article
C2 - 34017023
VL - 11
JO - Scientific Reports
JF - Scientific Reports
SN - 2045-2322
IS - 1
M1 - 10650
ER -