Skip to main navigation Skip to search Skip to main content

Feasibility and efficacy of a new non-invasive surface cooling device in post-resuscitation intensive care medicine

  • Moritz Haugk
  • , Fritz Sterz
  • , Martin Grassberger
  • , Thomas Uray
  • , Andreas Kliegel
  • , Andreas Janata
  • , Nina Richling
  • , Harald Herkner
  • , Anton N Laggner

Publications: Contribution to journalArticlePeer Reviewed

Abstract

AIM OF THE STUDY: There is sufficient evidence that therapeutic hypothermia after non-traumatic cardiac arrest improves neurological outcome and reduces mortality. Many different invasive and non-invasive cooling devices are currently available. Our purpose was to show the efficacy, safety and feasibility using a non-invasive cooling device to control patient temperature within a range of 33-37 degrees C.

MATERIALS AND METHODS: A convenience sample of patients who have been resuscitated successfully from cardiac arrest and were intended for mild hypothermia therapy according to the guidelines and inclusion criteria were studied in a prospective observational case series at an emergency department of a tertiary care university hospital. The Medivance Arctic Sun System provides a new, non-invasive approach to reach a target temperature of 33 degrees C quickly, to maintain the target temperature for 24h, and then to actively re-warm at 0.4 degrees C/h to normothermia. Cooling was applied using the Arctic Sun in 27 patients. Data are presented as median and the interquartile range (25, 75%).

RESULTS: Median age was 58 (49.5, 70) years. Time from cooling start to target temperature was 137 (96, 168)min, cooling rate was 1.2 degrees C/h (0.8, 1.5), stability of target temperature during hypothermia maintenance phase was satisfactory at 33.0 degrees C (32.9, 33.1), and duration of re-warming was 428 (394, 452)min.

CONCLUSION: Using the Arctic Sun System in post-resuscitation care medicine for cooling cardiac arrest survivors is feasible and has proven to be highly effective in lowering patients' temperature rapidly without inducing skin irritations.

Original languageEnglish
Pages (from-to)76-81
Number of pages6
JournalResuscitation
Volume75
Issue number1
DOIs
Publication statusPublished - Oct 2007

Austrian Fields of Science 2012

  • 302053 Emergency medicine

Keywords

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Critical Care
  • Equipment Design
  • Feasibility Studies
  • Heart Arrest/therapy
  • Humans
  • Hypothermia, Induced/instrumentation
  • Middle Aged
  • Prospective Studies
  • Resuscitation
  • Treatment Outcome

Fingerprint

Dive into the research topics of 'Feasibility and efficacy of a new non-invasive surface cooling device in post-resuscitation intensive care medicine'. Together they form a unique fingerprint.

Cite this